Test Object

What is the proper order for scan acquisition (scouts/topograms and series acquisitions)?

By |2021-04-21T17:24:32-05:00April 21st, 2021||

For the Standard, BMI dose-based protocols:  AP/PA scout at TLC: verify anteroposterior iso-centering and locate TLC scan endpoint  Lateral scout following TLC breath hold: verify right-to-left iso-centering, and select scan start point  TLC scan  Lateral scout following RV breath hold: locate RV scan endpoint  RV scan For the Reduced, dose

How do I fill the water bottle?

By |2021-04-21T17:16:58-05:00April 21st, 2021||

 To fill the water bottle:  Carefully fill the bottle with distilled water ensuring no bubbles are present.  If a bubble develops, please pour some of the water out of the water bottle.  Place your finger over the water bottle opening and carefully tilt upside down to alleviate any air bubbles.